Appendix C. PT4-L Agitator
The wiring for the agitator as controlled by this example program is as follows:
BLUE
WHITE
Agitator
+ -
12 V Battery
4 1
#7321 Relay
Crydom D1D07
3 2
Ground/G
Datalogger
Port C1
FIGURE C-3. Agitator Wiring
The following instructions would trigger the agitator as discussed in the
agitator manual.
The CR800, CR850, CR1000, CR3000, and CR5000 use the Portset()
instruction as follows:
CR1000 Portset() Instruction Example
Portset (1,1)
Delay (1,500,msec)
Portset (1,0)
The CR200(X)-series use the Portset() instruction as follows:
CR200(X) Portset() Instruction Example
Portset (1,1)
Delay(500,msec)
Portset(1,0)
C-3